IEC TR 61010-3:2003 pdf free download

IEC TR 61010-3:2003 pdf free download

IEC TR 61010-3:2003 pdf free download
IEC TR 61010-3:2003 pdf free download.Safety requirements for electrical equipment for measurement, control, and laboratory use
Clauses or subclauses which are not applicable need not be included in the prepared report, provided that these omissions are indicated in the contents list of the report. An example of a contents list based on IEC 61 01 0-1 : 2001 is given in Annex A for a report on a specific product.
4 Documents
The documents a) to d), together with any others required by the standard, should be reviewed and listed in the report:
a) general description of the equipment tested;
b) OPERATOR instructions;
c) installation instruction;
d) service instructions.
The documents e) to n), if applicable, should also be listed in the report:
e) component data sheets and certification details;
f) wire data sheets;
g) flammability test results or data sheets;
h) enclosure flammability test results and/or material data sheets;
i) comparative tracking indices of printed wiring board materials;
j) data sheets, test results or certification details for cathode-ray tubes;
k) constructional drawings and specifications for transformers including winding and insulation details;
l) test results for abnormal operation and fault conditions;
m) layout of printed wiring boards with primary and secondary hazardous voltages showing all working voltages on the tracks;
n) circuit diagrams, and assembly drawings of printed wiring boards.
5 Completion of the report
a) If additional information is required for any part of the report, it should be included on a separate sheet.
b) Clause 6 – Protection against electric shock – Block diagram of a system.
Where applicable, the table in Form A.5 should be completed and a block diagram included (where possible) to assist analysis of the product. See Figure 1 and Table 1 as examples of a block diagram and of a completed table. The letters A, B, etc., indicate points between which dielectric strength tests, or measurements of CREEPAGE DISTANCE or CLEARANCE , should be made.